Metal Ion Study 2002

Medium Term Serum Cobalt and Chromium Levels in Patients with Metal-metal Resurfacing
NOTE: This paper was presented at the New Trends of Joint Reconstructive Surgery 2002 Meeting, comparing serum levels in a control group, conventional THR, McKee Farrar, MOM Hip Resurfacing, and MOM THR (Sulzer), which concluded that metal ion levels are higher at 1 year with a small diameter MOM THR (Sulzer) than with resurfacing. As the Sulzer device has been in use for over 15 years with no associated problems reported to date, this is further evidence that this is currently not an issue.


V Reddy MD† J Dorairajan MD J Nevelos PhD* S Krikler PhD FRCS(Orth).
† University Hospitals Coven try & Warwickshire NHS Trust * Corin Medical†
Introduction
The conservative principles of hip resurfacing combined with very low wear metal-metal bearing technology may provide a very attractive solution for the younger arthritic hip. The long-term effects of the inevitable metal ion release however, remain a concern. Serum ion levels from metal-metal resurfacing hip patients were measured pre-operatively and then every year for up to five years.

Materials and Methods
All patients had had either a McMinn (1995-1996) or Cormet® (1997-) hip resurfacing (both manufactured by Corin Medical, UK) implanted by a single surgeon. Blood was collected using a standardised technique to ensure no metallic contamination. Chromium analysis was by graphite furnace atomic absorption (ETA-AAS) using a 4100ZL or A800 Perkin Elmer instrument. Cobalt was analysed by inductively coupled plasma mass spectrometry (ICP-MS). Measurements were taken from 39 patients with 7 of those having bilateral resurfacings. Pre-operative levels were only taken from 15 patients, and these were universally low (below 20nmol/l).

Results – General Population
In all patients the serum levels of cobalt and chromium increased following resurfacing although the extent of the increase varied greatly between patients. There was a general decrease in ion levels from ~2 years. The data shows a definite trend of decreasing ion levels after 3 years. This may be consistent with running-in wear and healing of the peri-prosthetic tissues providing a smooth, stable joint. The decline in serum level seems to start slightly earlier in younger patients (<40) than older patients (>60).
 

Discussion – General Population

  • Ion levels increase over the first two years before dropping to a generally low level.
  • Younger patients’ ion levels peak earlier than older patients – probably related to activity level.
  • There was no difference in ion levels between the McMinn and Cormet devices.
  • Bilateral resurfacings did not appear to elevate the ion levels significantly.
  • Insufficient sample numbers for statistical analysis.
Results – Outliers
The vast majority of patients studied had similar ion level release which remained below 250nl/l. There were, however three individual patients who had much higher ion levels which may cast light on possible causes of increased wear.

Case 1. Patient AP. 22 year old female at primary operation – Steep Acetabular Cup



Case 2. Patient LB. 42 year old female at primary operation Post operative Dislocation and Steep Cup



Case 3. Patient SM. 35 year old male at primary operation.






Discussion – Outliers
These cases were performed before the start of this study and so ion levels were not taken pre-operatively. Their levels were one or two orders of magnitude higher than the general population. (Note scale on the respective graphs). Also the ion levels show no sign of decreasing with time unlike the general population. Steep acetabular cup angles have been shown to be related to increased wear in hard-hard hip bearings (16) . Steep cups will also inevitably lead to an increased chance of dislocation (as in case 2). The large diameter heads used for hip resurfacing are naturally resistant to dislocation and Case 2 of this series remains the only dislocation to date in the senior author’s personal series of over 250 operations.

Conclusions
In the vast majority of patients with metal-metal hip resurfacing there is a small but early rise in serum metal ions over the first 2-3 years but the levels then gradually diminish over time .

Steep cups seem to be associated with increased levels, which could be due to wear and therefore increased ion release into the body.
 

READ COMPLETE STUDY

 

L10 Web Stats Reporter 3.15 LevelTen Hit Counter - Free PHP Web Analytics Script
LevelTen dallas web development firm - website design, flash, graphics & marketing
Contact Patricia

Advertisements

Advertise on
Surface Hippy

 

8000
Unique Visitors
per month use
Surface Hippy


Contact Patricia

Hip Talk generates over 50,000 page views a month.

Surface Hippy would like to Thank the many supporters of the website. 

The website content is not controlled by any medical companies, doctors or hospitals.

 

What's New?

Dr. Bose presented plaque by overseas patients

Dr. Bob Arnot's Bi-lat hip resurfacing story with Dr. Su

What hip resurfacing device is most popular?

Hip Resurfacing Surgeons Video Interviews

Web design by Patricia Walter Copyright Surface Hippy 2006

Statistics Page

Mission Statement - Surface Hippy is a patient to patient guide to hip resurfacing. It does not provide medical advice. It is designed to support, not to replace, the relationship between patient and clinician.
Advertising - Revenue from this site is derived from Google Ads, commercial advertising and individual donations.
Privacy - Surface Hippy does not share email addresses or personal information with any group or organization.
Content - Surface Hippy is not controlled or influenced by any medical companies, doctors or hospitals.
All content is controlled by Patricia Walter  -
Joint Health Sites  LLC