The Influence of Head Diameter, Clearance, Cup Position, and Head Position on Wear Rates in Metal-on-Metal Resurfacing

"The results also confirm clinical ion level measurements that steep cup angles can substantially increases wear"

Updated 1/8/09

Presented at the 2nd Annual Total Hip Resurfacing Arthroplasty Course in LA Oct. 2008

By: John Fisher
Co-Authors: Ian Leslie, Sophie Williams, Eileen Ingham, Graham Isaac
Institute of Medical and Biological Engineering
University of Leeds

Background: There are considerable variations in metal ion levels and metallic wear rates in patients with metal on metal resurfacing. In this in vitro study the effect of design variables of head diameter and bearing clearance and patient variables of cup and head position on metal ion levels and wear rates are were investigated. Methods: Hip joint simulator studies were carried out on size 39mm and size 55mm metal on metal resurfacing with the same design. Size 55mm diameter bearings with 110 micrometer diametrical clearance were compared to size 54mm diameter with larger >250 micrometer diametrical clearance. The wear rates of size 39mm bearings with a standard cup position of 45 degrees, were compared to a steep cup position of 60 degrees and to a steep cup position combined with micro separation associated with head offset deficiency.

Results: Initial bedding in wear rates and ion levels were higher with size 39mm bearings compared to size 55mm, but in long term after 15 million cycles there was no difference in the steady state wear rates. Bearings with the larger diametrical clearance had higher initial wear and steady state wear rates at five million cycles. Cup position and head position resulted in much greater increases in wear. For the 39 mm bearings, a 60 degree cup position resulted in a 9 fold increase in wear. A steep cup and microseparation resulted in a 17 fold increase in wear after two million cycles and a 39 fold increase in wear compared to the long term steady state wear rate.

Conclusions: The study confirmed clinical ion level studies of increased wear with larger clearance bearings. The results also confirm clinical ion level measurements that steep cup angles can substantially increases wear. The study also indicates that offset deficiency and microseparation may be responsible for extremely high wear rates and ion levels found in some retrievals and some patients. Further work is needed to investigate effect of different head sizes with steep cups and microseparation and the effect of version.

Advertisement


View My Stats

Statistics Page

Web design by Patricia Walter Copyright Surface Hippy 12/11/2005

Mission Statement - Surface Hippy is a patient to patient guide to hip resurfacing. It does not provide medical advice. It is designed to support, not to replace, the relationship between patient and clinician.
Advertising - Revenue from this site is derived from commercial advertising and individual donations.
Any advertisement is distinguished by the word "advertisement"
Privacy - Surface Hippy does not share email addresses or personal information with any group or organization.
Content - Surface Hippy is not controlled or influenced by any medical companies, doctors or hospitals.
All content is controlled by Patricia Walter  -
Joint Health Sites  LLC
This site is published by Joint Health Sites LLC, which is solely responsible for its content.
The advertisements on this site are not intended by the advertisers as an endorsement of the site's content.
The advertisers shall not be liable for any errors or omissions in the site's content,
nor liable for any damages from any person's actions based in reliance on the site's content.